Abstract
To solve extended acquisition time issues inherent in the conventional hopping-scanning mode of scanning ion-conductance microscopy (SICM), a new transverse-fast scanning mode (TFSM) is proposed. Because the transverse motion in SICM is not the detection direction and therefore presents no collision problem, it has the ability to move at high speed. In TSFM, the SICM probe gradually descends in the vertical/detection direction and rapidly scans in the transverse/nondetection direction. Further, the highest point that decides the hopping height of each scanning line can be quickly obtained. In conventional hopping mode, however, the hopping height is artificially set without a priori knowledge and is typically very large. Consequently, TFSM greatly improves the scanning speed of the SICM imaging system by effectively reducing the hopping height of each pixel. This study verifies the feasibility of this novel scanning method via theoretical analysis and experimental study, and compares the speed and quality of the scanning images obtained in the TFSM with that of the conventional hopping mode. The experimental results indicate that the TFSM method has a faster scanning speed than other SICM scanning methods while maintaining the quality of the images. Therefore, TFSM provides the possibility to quickly obtain high-resolution three-dimensional topographical images of extremely complex samples.
Introduction
Using a scanning probe microscope to achieve three-dimensional topographic imaging of living biological cells and dynamic observation has been a long-time challenge, but the advent of the scanning ion-conductance microscope (SICM) has provided an important development thereto. As a potential new member of the scanning probe microscope techniques, SICM was first proposed and invented by Hansma et al. in 1989. Because SICM uses the detected ion current flowing through the tip of its micropipette to control the distance between its probe and the sample surface, it can realize noncontact, high-resolution imaging for biological samples in a physiological environment (Korchev et al., 1997; Shevchuk et al., 2006). In recent years, SICM has been widely used in quantitative transmission of nanoparticles (Bruckbauer et al., 2002; Babakinejad et al., 2013; Ivanov et al., 2015), in drug reactions of biological micro-organizations (Yang et al., 2012) and in electrochemical detection (Takahashi et al., 2010; Nadappuram et al., 2013; O’Connell et al., 2014; Şen et al., 2015). It thus has widespread and promising applications in micro- and nanofabrication, materials research and development, stoichiological research and drug development.
Hansma et al. (1989) further proposed a direct current (DC) scanning mode, which maintains a constant ion current during the scanning process by controlling the z-direction motion of the micropipette. However, the ion current is inevitably affected by ionic current drift in the DC scanning process, resulting in collisions between the tip of the micropipette and the sample surface and thereby reducing the imaging stability.
To improve the SICM imaging stability and accuracy and its capability for imaging complex surface samples, various research groups have developed several important imaging modes. Pastre et al. (2001) and Shevchuk et al. (2001) have developed a distance modulation scanning mode, named the alternating current (AC) mode. In this mode, the SICM images using a probe that vibrates with an amplitude of a few dozen nanometers, where the amplitude of the AC ionic current is detected using a lock-in amplifier locked to the vibration frequency of the probe (Proksch et al., 1996; Pastre et al., 2001). This AC mode can overcome the effects of ionic current drift and improve the imaging stability and the sensitivity to distance measurements, but the vibration frequency of the probe restricts its scanning speed. McKelvey et al. (2014) and Li et al. (2014) have separately proposed another modulation scanning mode that applies an oscillating bias between a quasi-reference counter electrode (QRCE) in the probe and a second QRCE in the bulk solution to eliminate any physical oscillation of the probe, thereby generating an oscillating ion current feedback signal. Mann et al. (2002) have developed the backstep mode and after implementing further refinements, Novak et al. (2009) have developed the hopping mode to further improve the SICM imaging stability for biological samples with complex topographies. The hopping mode greatly enhances the SICM imaging capability and stability, but it inevitably reduces the scanning speed of the probe. Zhukov et al. (2012) have proposed a hybrid scanning mode for fast SICM that quickly scans on samples with relatively flat surfaces and, to some extent, compensates for the hopping mode. Zhuang et al. (2017) have proposed a new scanning mode by utilizing the pipette predicted movement in the horizontal direction. In this method, the pipette parameters, such as the half cone angle, the ratio of the inner to outer radius, and the opening radii of the pipette tip, play a critical role in anticipating the upcoming raised topography in the horizontal direction. To achieve a maximal detectable distance, it is necessary to balance the relationships between the above-mentioned parameters. In addition, it is also necessary to balance the relationship between signal-to-noise ratios of the ionic current and the feedback threshold (Zhuang et al., 2017).
Hitherto, the majority of the existing SICM systems have adopted the conventional hopping mode, which takes a long time to image samples with complex topographies. This paper proposes a new scanning method to improve on the scanning speed of the traditional hopping mode without sacrificing its powerful detection capability. First, a home-built SICM system is developed that can operate in a new transverse-fast scanning mode (TFSM). Then, we report on contrast experiments carried out with TFSM on polydimethylsiloxane (PDMS) samples, human breast cancer cells and hippocampal neuronic cells with complex morphologies. Finally, the imaging speed and quality of SICM systems operating in the traditional hopping mode and in TFSM are compared with determine whether TFSM exhibits a better scanning speed without reducing the imaging quality.
Theoretical Analysis
Principles of the Method
The imaging speed of the conventional hopping mode of SICM is slow, mainly owing to the following two aspects: first, owing to the influence of the movement inertia, the probe continues to move in the original direction for a distance after reaching the target location. As the probe speed is increased, the influence of the movement inertia will also increase. Because the probe speed in the z-direction directly affects the SICM imaging quality, the probe velocity in this direction is seriously limited. Second, in the conventional hopping mode it is common to set an excessive hopping height because the highest level for the scanning area of the sample is unknown. However, an excessive hopping height significantly increases the imaging time of the probe.
Schematic diagram of proposed scanning mode. a: The illustration of hopping height hm corresponding to the m-transection of the sample (M×N, M and N is the numbers of rows and columns, respectively). b: The obtained hopping heights of the whole sample using horizontal fast scanning method. c–e: The process of detecting the highest point of m-line in horizontal fast scanning method. f: The scanning process of m-line in transverse fast scanning mode (TFSM) with automatically detecting the highest point. g: The scanning process of m-line in conventional hopping mode with the artificially setting hopping height.
The key to implementing TFSM is to control the probe so that it can quickly and reliably detect each highest point of the scanning line. This work presents a new method to rapidly obtain the highest points of each scanning line, as described below. Figures 1c-1e depict the detailed process of detecting the highest point of a scanning line in a sample with the TFSM, which is labeled as the m-line in Figure 1a. First, the probe moves toward the sample surface at a speed vz along the z-direction, moving a distance Δz. Then, the sample is driven by an x-directional piezoelectric actuator to quickly move in the x-direction at a speed vx. During this procedure, if the ionic current, i, of the SICM does not quickly decrease to a set value iset, the probe will again move Δz in the z-direction and the sample will move in the x-direction at a speed −vx. This procedure is performed repeatedly until i decreases to iset, at which point the height value of line m is set at (Fig. 1e). Based on the above-mentioned process, the SICM system has completed the detection of the highest point of the m-line.
Theoretical Comparison of the Conventional Hopping Mode and TFSM
Equation (3) shows that a smaller and a larger Δz reduce the time required in the TFSM compared with that of the conventional hopping mode. Novak et al. (2014) have measured the dynamic interaction between a nanoparticle and a living cell with vz up to 500 nm/ms by utilizing the double z-piezo structure that can rapidly withdraw the SICM probe. The single z-piezo is usually employed in the conventional hopping mode and the value of vz is typically 50–200 nm/ms. In the literature, Jung et al. (2015) reported that the image artifacts began to appear on the surface of the sample as a result of the reduced stability when approach rate is 300 nm/ms. Watanabe & Ando (2017) have developed a high-speed XYZ-nanopositioner with vertical travel range of ~6 μm and the tip approach rate is 400 nm/ms. To simplify the comparison of imaging time for the m-line of the sample, typical setting parameter values are input into equation (3). Next, we assume vz=200 nm/ms, vx=4 mm/s, Δz=100 nm, Δx=500 nm, N=100, h0=5 μm, and
. With these values we obtain a ΔTm=0.9375s>0, which clearly shows that the m-line imaging time of the TFSM is less than that of the conventional hopping mode.
Theoretical Comparison of the Standing Approach (STA) Mode and TFSM
Comparison of standing approach (STA) mode and transverse fast scanning mode (TFSM) for imaging the m-line of the sample. a,c: The scanning trajectories using STA mode for the flat and steep samples, respectively. b,d: The scanning trajectories using TFSM for the flat and steep samples, respectively.
From equation (7), it can be seen that hmw is determined by several scanning parameters and sample parameters. Based on equation (7), we can estimate the value of hmw and further compare the imaging speed of the STA mode and TFSM.
Parameter Settings of the TFSM Smart Hopping Mode
a: Approach curve of the micropipette. b: Scanning electron microscope image of the micropipette tip.
Theoretically, the probe detects the sample when the ion current begins to decrease. However, there are many kinds of experimental noise that lead to obtaining a false highest point. Therefore, we considered that the probe detected the highest point when the ion current decreased by 1% in actual experiments. As shown in Figure 3a, the distance between the tip of the micropipette and the sample surface was about 228 nm when the ion current decreased to the set point. In this paper, considering the detection speed and reliability, we set the approaching step Δz to 100 nm. Taking into account the influence of experimental environmental noise and the movement inertia of probe, the hopping height for each line was set as the sum of the height of the peak point of that line and the inner diameter of the micropipette opening, to improve the scanning speed without affecting the SICM imaging quality. As shown in the SEM image in Figure 3b, the probe opening radius is about 130 nm.
Instrumentation and Materials
Instrumentation
In the experiments, we used a home-built SICM system mainly composed of piezoelectric ceramic and a piezo-driver (Physik Instrument, Germany), Ag/AgCl electrodes, a nano-microprobe, 16-bit DA and AD modules, a patch clamp amplifier, core control chip (Xilinx FPGA), a host personal computer and a serial communication module.
Pipettes
The SICM probes were selected from borosilicate capillaries (Sutter Instrument Company, Novato, CA, USA) with internal diameters of 0.58 mm, external diameters of 1 mm, and lengths of 10 cm. The borosilicate capillaries were pulled using a laser pipette puller (P-2000, Sutter Instrument Company), whereby different opening radii were obtained by adjusting the puller parameters. The opening radii of the probes were about 130 nm, which were used in sets of experiments.
Samples
A series of contrast experiments were carried out on PDMS samples and living cell samples. One PDMS sample featured cylinders (4 μm radius, 2.5 μm height) and the other PDMS sample exhibited six-pointed stars (20 μm circumscribed circle radius, 2.5 μm height). All the PDMS samples are fabricated using imprint lithographic methods. The living cell samples included cells from the human breast cancer cell lines MCF-7, MDA-MB-231, murine cardiomyocytes and hippocampal neuronic cell with complex morphology. The PDMS and living cell experiments, respectively, used a KCl (0.1 mol/L) solution and a cell buffer solution during scanning. The cell buffer solution was prepared by the following process: first, a certain mass of NaCl, Na2HPO4, and NaH2PO4 were dissolved in ultrapure water, and the pH value of phosphate solution was then adjusted to 7.2 using an HCl (1 mol/L) solution. Finally, the buffer solution was filtered with filter paper (filtration pore 20 nm) and was sterilized.
Results and Discussion
Quantitative Evaluation Methods
Comparison of Imaging Speed and Quality with Conventional Hopping Mode
The comparison experiments for the imaging speed were conducted by scanning PDMS samples of known height and scanning human breast cancer cell samples of unknown height. We scanned identical sample areas ten times using the same micropipette with the same scanning speed in the two different scanning modes in the self-built SICM system. The tip–sample distance during scanning greatly influenced the imaging quality (Thatenhorst et al., 2014; Rheinlaender & Schäffer, 2015), so we scanned the comparison experiment samples with a 1.0% set point (point “A” in Fig. 3a).
Comparison of imaging speeds and qualities for the polydimethylsiloxane (PDMS) samples with known height in transverse fast scanning mode (TFSM) and conventional hopping mode. a: Scanning image of PDMS 1 in TFSM. b: Scanning image of PDMS 1 in conventional hopping mode. c: Scanning image of PDMS 2 in TFSM. d: Scanning image of PDMS 2 in conventional hopping mode. e: Comparison of the average pixel imaging frequency for PDMS 1 and PDMS 2 in two scanning modes. f: Comparison of mean squared of scanning images height fluctuation of PDMS 1 and PDMS 2 in two scanning modes.
Figures 4e and 4f plot the average pixel imaging frequency and the mean squared error (MSE) of the height fluctuation, respectively, in the images obtained by the TFSM and the conventional hopping mode. The calculation results show that the average pixel imaging frequencies of the PDMS samples 1 and 2 are 21.46 and 23.19 Hz, respectively, in conventional hopping mode; while that in the TFSM are about 38.24 and 39.19 Hz, respectively (Fig. 4e). In TFSM, the pixel imaging frequencies of samples 1 and 2 are increased by 78.2 and 69.0%, respectively, from that of the conventional hopping mode. As shown in Figure 4f, the MSEs of imaging the PDMS samples 1 and 2 in the conventional hopping mode is 398.56 and 425.66 nm2, respectively; while that in the TFSM are 406.82 and 410.23 nm2, respectively.
Comparison of imaging speeds and qualities for the human breast cancer cell samples with unknown height in transverse fast scanning mode (TFSM) and conventional hopping mode. a,c: The images of cell 1 (MDA-MB-231) and cell 2 (MFC-7) in TFSM, respectively. b,d: The images of cell 1 and cell 2 in conventional hopping mode, respectively. e: Comparison of average pixel imaging frequency for cell 1 and cell 2 in the two scanning modes. f: Comparison of mean squared of scanning images height fluctuation of cell 1 and cell 2 in two scanning modes.
The experimental results indicate that the average pixel imaging frequency of the TFSM is larger than that of the conventional hopping mode, and the MSEs in the two different scanning modes are approximately equal, whether imaging PDMS or living cells. Furthermore, the imaging speed for samples possessing unknown heights (i.e., living cell samples) is vastly improved over that of samples with known heights (i.e., PDMS samples).
The aim of this work is to optimize the scanning parameters and improve the imaging rate in the conventional hopping mode SICM. For imaging the samples with unknown topographies (such as cells, metals, and nonmetallic samples), the surface topographies of samples may be flat or extremely complex and even the locations are unknown. Therefore, the prescanning method (Zhukov et al., 2012) is usually adapted to obtain some knowledge of the sample morphology before the subsequent scan can be planned. Unfortunately, the prescanning method may not detect local abrupt sample structures and also reduce the scanning stability and imaging quality owning to the probe collision. Moreover, prescanning itself is also a very time-consuming task. Accordingly, to ensure scanning stability and imaging quality, a larger hopping height is usually employed in hopping mode SICM.
Comparison of imaging speeds and qualities for the hippocampal neuronic cell samples with unknown height in transverse fast scanning mode (TFSM) and conventional hopping mode. a: Imaging result of hippocampal neuronic cells in TFSM. b: Imaging result of hippocampal neuronic cells in conventional hopping mode. c: Comparison of average pixel imaging frequency for hippocampal neuronic cells in the two scanning modes. d: Comparison of mean squared of scanning images height fluctuation of hippocampal neuronic cells in two scanning modes.
Maximum Slopes of All Samples and the Corresponding Critical Amplitudes in Standing Approach (STA) Mode.
| Samples . | Maximum Slopes (dz/dx) . | STA Mode [Critical Amplitudes (μm)] . |
|---|---|---|
| PDMS1 | 71.34° (0.296/0.1) | ~1.61 |
| PDMS2 | 59.36° (0.843/0.5) | ~1.52 |
| Cell1 (MDA-MB-231) | 48.24° (0.56/0.5) | ~1.75 |
| Cell2 (MFC-7) | 40.70° (0.43/0.5) | ~1.66 |
| Hippocampal neuronic cells | 68.18° (1.249/0.5) | ~4.65 |
Comparison of Imaging Speed (TFSM and Hopping Mode) with Different Hopping Heights
Comparison of imaging speeds for the cell samples A (murine cardiomyocytes) and B (MFC-7) with different hopping heights in transverse fast-scanning mode (TFSM) and conventional hopping mode. a,c: The images of cell A and B in TFSM, respectively. b,d: The images of cell A and B in conventional hopping mode, respectively. e: Comparison of average pixel imaging frequency for cell A with different hopping heights in two scanning modes. f: Comparison of average pixel imaging frequency for cell B with different hopping heights in two scanning modes.
| Samples . | Methods . | Average Pixel Image Frequencies (Hz) [Hopping Height (μm)] . | |||
|---|---|---|---|---|---|
| Cell A | TFSM | 32.015 (8 μm) | 33.871 (7 μm) | 34.856 (6 μm) | 35.796 (5 μm) |
| Hopping mode | 20.501 (8 μm) | 23.482 (7 μm) | 25.265 (6 μm) | 27.671 (5 μm) | |
| Cell B | TFSM | 19.269 (20 μm) | 20.67 (18 μm) | 21.278 (16 μm) | 22.056 (15 μm) |
| Hopping mode | 11.053 (20 μm) | 14.048 (18 μm) | 15.565 (16 μm) | 17.078 (15 μm) | |
TFSM, transverse fast-scanning mode.
Comparison of Imaging Speed and Stability with Previous Method (Zhuang et al., 2017)
In this part, considering the previous work (Zhuang et al., 2017) also used horizontal scanning to detect a change in the ion current to modify the hopping paths. First, we demonstrate the main differences of the two methods, and then compare their imaging speed and stability with PDMS samples and cell samples.
In TFSM, the purpose of the horizontal scanning is that the pipette can quickly detect the highest point of each scanning line. In the previous method, the horizontal scanning is a predicted movement for upcoming raised topography in the next measurement point. TFSM mainly employs the tip detection capability in the vertical direction and the opening radius of the tip is the main parameter. While the previous work employs the tip detection capability in the sidewall direction and the half cone angle, the ratio of the inner to outer radius and the opening radii of the pipette tip are the main parameters.
Comparison of imaging speeds and success rates of scanning for polydimethylsiloxane (PDMS) samples with different topographies in transverse fast-scanning mode (TFSM) and previous method. a,c: The images of PDMS A and PDMS B in TFSM, respectively. b,d: The images of PDMS A and PDMS B in previous method, respectively. e: Comparison of average pixel imaging frequency in two scanning methods. f,g: Comparison of success rates of scanning for PDMS A and PDMS B in two methods with different measurement points, respectively.
Comparison of imaging speeds and success rates of scanning for cell samples with different topographies in transverse fast-scanning mode (TFSM) and previous method. a,c: The images of cell A (murine cardiomyocytes) and cell B (MFC-7) in TFSM, respectively. b,d: The images of cell A and cell B in previous method, respectively. e: Comparison of average pixel imaging frequency in the two scanning methods. f,g: Comparison of success rates of scanning for cell A and cell B in two methods with different measurement points, respectively.
Success Rates of Scanning with Different Number of Pixels.
| . | . | Success Rates of Scanning (%) (Number of Pixels) . | ||
|---|---|---|---|---|
| Samples . | Methods . | % (32×32) . | % (100×100) . | % (256×256) . |
| PDMS A | TFSM (proposed method) | 98.3 | 98.3 | 96.6 |
| Previous method (Zhuang et al., 2017) | 95.0 | 91. 6 | 86.6 | |
| PDMS B | TFSM | 98.3 | 98.3 | 96.6 |
| Previous method | 96.7 | 88.4 | 84.2 | |
| Cell A | TFSM | 96.6 | 91.6 | 88.3 |
| Previous method | 90.0 | 35.0 | 15.0 | |
| Cell B | TFSM | 95 | 91.7 | 88.33 |
| Previous method | 88.3 | 45 | 21.6 | |
PDMS, polydimethylsiloxane; TFSM, transverse fast-scanning mode.
As seen from the Table 3, the imaging speed of the previous method is larger than that of TFSM only when the scanning process of the previous method has not failed. However, the success rate of the previous method is lower than that of TFSM whether scanning PDMS or living cell samples. This is mainly because there are more requirements for detecting the changes in ion current in the process of predicting the upcoming raised topographies in the previous method, especially for the cell samples. The increasing interference factors that occur due to the physiological activities between cells in the cell environment affect the detecting process and finally lead to the decrease of predicting precision and the failure of scan.
Influence of Scanning Direction on Imaging Speed of TFSM
The influence of scanning direction on imaging speed in the transverse fast-scanning mode (TFSM). a: 0° direction; (b) 45° direction; (c) 90° direction; (d) comparison of average pixel imaging frequency for cell 1 (MDA-MB-231) from different scanning directions using TFSM.
Conclusions
This paper presents the TFSM of SICM to improve upon the imaging speed of the conventional hopping mode, where the feasibility and advantages of the TFSM are validated by both theoretical analysis and experimental studies. The TFSM can attain fast imaging by the setting of a reasonable hopping height for each scanning line that depends on the height of the highest point in each line. Using a home-built SICM system, a series of comparative experiments were carried out on PDMS samples with known feature heights, human breast cancer cells with unknown feature heights and hippocampal neuronic cells with complex topographies, using both the TFSM and conventional hopping mode. By comparing the imaging speed and imaging quality of the two different scanning modes, we show that the imaging speed of TFSM is greater, while the imaging quality of the two different modes are nearly identical. Moreover, further studies on the characteristics of the TFSM show that the imaging speed is related to the scanning direction of the probe.
In summary, herein we introduce a new scanning mode for fast SICM, and also create a new opportunity to realize fast imaging for samples with extremely complex morphologies.
Acknowledgments
The authors thank the National Natural Science Foundation of China (Project no. 51375363) and Industrial Research Project of Science and Technology Department of Shaanxi Province (Project no. 2013GY2-04) for funding this work. In addition, the authors are grateful for the permission to use the human breast cancer cell samples from the Bioinspired Engineering and Biomechanics Center at Xi’an Jiaotong University. The authors thank Dr. Wang Changhe from the School of Life Science and Technology in Xi’an Jiaotong University for providing the hippocampal neuron cell samples in this work.
References
Author notes
Cite this article: Zhuang J, Wang Z, Li Z, Liang P, Vincent M (2018) Smart Scanning Ion-Conductance Microscopy Imaging Technique Using Horizontal Fast Scanning Method. Microsc Microanal 24(3): 264–276. doi: 10.1017/S1431927618000375